Monitor the problem to see if it is always initiated by the same module (HPLC LAN card suspect) or if it varies (Computer LAN suspect). You could start by updating the computer's LAN firmware to see if that helps, otherwise a new computer may be needed. In that case, a new computer LAN card (or motherboard, as most are integrated) is required. This same problem can also be caused by the computer's LAN board (which communicates to the HPLC's LAN board) dropping the signal too. That seems more likely due to the LAN board in the detector being "flaky". In your case, the communication drop probably initiated from a DAD daisy chained to a TCC. Communication which is lost to just one or two modules, can indeed be from a defective $5 Ethernet cable (CAN cable) so it is always worth replacing them to start with. That is consistent when it causes a Sequence to abort during a run, but continue to run the system.
The initial errors indicate a lose of communication to the modules over the instrument network. Have you applied the two available service packs to your B.04.03 software? I'm currently rinsing the flow cell, but any other help or information would be helpful. The errors encountered are:Įrror 41300 Occurred (invalid path, Agilent claims this was fixed with an update)Įrror 10010, can't write to DAD1.UV- This error results from lost spectra during acquisition. I restarted the sequence and tried to finish my batch, however I encountered another set of errors.
Any help or insight would be greatly appreciated. I plan to replace the CAN cables attached to the modules mentioned, but I was wondering if anyone else had any experience encountering this error. I've searched online and haven't found any answers to resolve this issue. G1315 (Detector) "Starting analysis aborted via CAN" G1316A (Column Compartment) "Starting analysis aborted via CAN" Everything runs fine, but recently I've been getting an error that causes a shutdown and doesn't provide a lot of information as to what caused the error.
AGILENT CHEMSTATION HPLC ABORT SERIES
I am currently running an Agilent 1100 Series HPLC with Chemstation Rev.